Warning
The fume hood must be running whenever liquid resin or isopropanol is handled. See the safety instructions of the Resin module.
Ductless chemical fume hood with activated carbon filter
ChemFAST Elite 15 fume hood
Ductless (filtering) chemical fume hood that protects against chemical vapours and aerosols, used in the resin room.
A fume hood (here a ductless, filtering one) draws air in through its front working opening, passes it through a prefilter and then an activated carbon filter that traps chemical vapours and aerosols, before recirculating it into the room. It protects the operator and the environment from the fumes, vapours and odours released when handling chemicals.
In the resin room, it provides additional protection when using liquid resin and isopropanol (cleaning, post-processing).
It is not equipment for large quantities of evaporated solvents, acid digestions, or chemicals of unknown composition.
| Parameter | Value |
|---|---|
| Brand / model | FASTER ChemFAST Elite 15 |
| Overall dimensions (W x H x D) | 1500 x 1120 x 760 mm |
| Usable dimensions (W x H x D) | 1438 x 660 x 600 mm |
| Working opening (sash lowered) | 200 mm |
| Maximum opening | 455 mm |
| Weight | 115 kg |
| Supply | 230V AC, 50Hz |
| Power consumption | 210 W (measured at 0.55 m/s) |
| Noise level | ~52 dB(A) (46 dBA measured on delivery) |
| Filtration | Prefilter + activated carbon filter (type chosen to suit the product) |
| Control | Microprocessor: automatic airflow regulation, safety alarms (Elite version) |
Start-up: switch on the lighting, check that the front sash is lowered, start the unit. A stabilisation phase of about 40 seconds follows (intermittent audible alarm): do not handle anything under the hood during that time. Once the alarm stops, the fume hood is ready for use.
During use: always work with the front sash fully lowered. That is what guarantees the maximum face velocity. A wide-open sash greatly reduces the effectiveness of the protection.
Shutdown: remove the equipment from the work area, clean the work surface, switch off the lighting, then switch off the unit.
A performance check (face velocity, airflow, absence of dead zones) is recommended once a year.